| BioSpectrometer® kinetic |
Absorption measuring range | 0 A – 3.0 A (260 nm) |
BSA concentration range | 7.5 ng/µL – 45.45 µg/µL |
BSA concentration range (UV 280 nm) µCuvette G1.0 | 75 ng/µL – 45.45 µg/µL ±0.03 µg/µL at A = 0 ±0.076 µg/µL at A = 1 (0.5 %) |
BSA concentration range (UV 280 nm) UVette 10 mm | 7.5 ng/µL – 4.5 µg/µL ±0.003 µg/µL at A = 0 ±0.007 µg/µL at A = 1 (0.5 %) |
BSA concentration range (UV 280 nm) UVette 2 mm | 37.5 ng/µL – 22.5 µg/µL ±0.015 µg/µL at A = 0 ±0.038 µg/µL at A = 1 (0.5 %) |
Beam receiver absorption | CMOS photodiode array |
Cuvette shaft | 12.5 mm x 12.5mm |
Emission wavelength I fluorescence | – |
Emission wavelength II fluorescence | – |
Excitation wavelength fluorescence | – |
Fluorescence measuring range | – |
Light path height | 8.5mm |
Light source absorption | Xenon flash lamp |
Light source fluorescence | – |
Measuring principle absorption | Single-beam absorption spectrophotometer with reference beam |
Measuring principle fluorescence | – |
Method memory | >100 method programs |
Operator guidance language | Spanish, Italian, French, English, German |
Power consumption | Approx. 15W during operating step Approx. 5W with dimmed display |
Random error absorption | ≤0.002 at A=0 ≤0.005 (0.5%) at A=1 |
Random error fluorescence | – |
Random error wavelength | ±1 nm |
Random temperature error | – |
Smallest step size | 1 nm |
Spectral bandwidth | ≤4 nm |
Systematic error absorption | ±1 % (A = 1) |
Systematic error wavelength | ≤0.5 nm |
Systematic temperature error | – |
Wavelength range | 200nm–830nm |
Wavelength range absorption | Scan (nm): 200 – 830 at 1 nm increments |
dsDNA concentration range | 2.5 ng/µL – 1,500 ng/µL |
dsDNA concentration range (UV 260 nm) UVette 10 mm | 2.5 ng/µL – 150 ng/µL ±0.1 ng/µL at A = 0 ±0.25 ng/µL at A = 1 (0.5 %) |
dsDNA concentration range (UV 260 nm) UVette 2 mm | 12.5 ng/µL – 750 ng/µL ±0.5 ng/µL at A = 0 ±1.25 ng/µL at A = 1 (0.5 %) |
dsDNA concentration range (UV 260 nm) µCuvette G1.0 | 25 ng/µL – 1,500 ng/µL ±1 ng/µL at A = 0 ±2.5 ng/µL at A = 1 (0.5 %) |
Interfaces | › USB master: for USB stick and thermal printer DPU-S445 › USB slave for connecting to a PC › Serial interface RS-232: for thermal printer DPU-414 › Ethernet interface RJ45: For connecting to a network printer |
Power supply | 100–120V, 50–60 Hz |
Dimensions (W?x?D?x?H) | 295x 400x 150mm/ 11.6x 15.7x 6" |
Weight w/o accessories | 5.4 kg |
Interval between two measuring points | – |
Measuring time range | – |
Adjustable temperature range | – |